FDA announces Class I recall of 4,800 imaging catheters due to risk of vascular injury

Abbott Medical is recalling multiple lots of its Dragonfly OpStar optical coherence tomography (OCT) intravascular imaging catheter due to issues with one its radiopaque marker bands becoming loose, according to an advisory from the U.S. Food and Drug Administration (FDA).

The FDA is categorizing this as a Class I recall, meaning the use of these devices could lead to serious injuries or death. Five incidents and one injury have been reported so far. In two instances, the marker band in question, the proximal marker, separated completely from the device while it was being used.

“A loose marker band that has separated from the device may remain in the body after the catheter is removed, potentially leading to vascular injuries, including but not limited to embolism , thrombosis, dissection, ischemia, infarction, infection, or death,” according to the advisory.

The Dragonfly OpStar Imaging Catheter is designed to provide OCT inside the coronary arteries of patients with suspected coronary artery disease.

Michael Walter
Michael Walter, Managing Editor

Michael has more than 16 years of experience as a professional writer and editor. He has written at length about cardiology, radiology, artificial intelligence and other key healthcare topics.

Around the web

Eleven medical societies have signed on to a consensus statement aimed at standardizing imaging for suspected cardiovascular infections.

Kate Hanneman, MD, explains why many vendors and hospitals want to lower radiology's impact on the environment. "Taking steps to reduce the carbon footprint in healthcare isn’t just an opportunity," she said. "It’s also a responsibility."

Philips introduced a new CT system at ECR aimed at the rapidly growing cardiac CT market, incorporating numerous AI features to optimize workflow and image quality.

Trimed Popup
Trimed Popup